InertSep SPE cartridges from GL Sciences are known for their high purity and high reproducibility. The selection of these cartridges extends to various materials, modifications and formats. In addition to polymers and silica, carbon, aluminium oxide and magnesium silicate are also available as materials. In addition, modifications are available for the normal and reverse phase as well as for ion exchange and special applications. The cartridges are available in Luer lock connections (Slim-J and Slim), standard cartridges and large size cartridges (LSC).
InertSep WAX FF/GCB cartridges are new cartridges designed specifically for EPA Method 1633. These cartridges replace the use of two SPE cartridges connected in series.
Currently and until 30.09.2024 you will receive a 35% introductory discount on the Inertsep WAX FF/GCb and InertSep GCB/WAX FF SPE cartridges. You can find more information about these cartridges here.
InertSep WAX FF/GCB and InertSep GCB/WAX FF are new SPE cartridges that have been specially developed for the EPA method 1633 for analysing PFAS in wastewater, surface water, groundwater, soil, sewage sludge, sediment, landfill leachate and fish tissue. For water samples, the method provides for extraction with WAX first and then with GCB. For soil samples, the order is reversed and extraction with GCB takes place before WAX. The two-layer system of the new cartridges fulfils the requirements of EPA Method 1633.
EPA method 1633 for the determination of PFAS components.
Column: InertSustain AQ-C18 1.9µm, 50x2.1mm (5020-89938)
Delay Column: Delay Column for PFAS 30x3.0mm (5020-90005)
Mobile Phase A: Methanol
Mobile Phase B: 20mmol/l Ammoniumacetate in Water
Gradient:
Time / min | %A | %B |
0 | 5 | 95 |
0.5 | 5 | 95 |
3.0 | 40 | 60 |
16.0 | 80 | 20 |
18.0 | 80 | 20 |
20.0 | 95 | 5 |
22.0 | 95 | 5 |
25.0 | 5 | 95 |
System: Exion HPLC System (SCIEX); QTRAP 650+ LC-MS/MS System (SCIEX)