HALO PFAS and HALO PFAS Delay columns have been developed specifically for the analysis of Per- and polyfluoroalkyl substances (PFAS). The densely bonded, extensively endcapped ODS stationary phase of HALO PFAS provides an application assured and method qualified solution for PFAS analysis. Every HALO® PFAS lot is quality assurance tested by LC-MS/MS analysis of a mix of 17 PFAS consisting of a broad range of short and long chain compounds.
Phase | Modification | Particle size / μm | Pore size / Å | End-capping | Surface area | pH range | Tmax |
HALO® PFAS | C18 | 2.7 µm | 90 Å | + | 135m2/g | 2.0-9.0 | 60 °C (a) 40 °C (b) |
HALO® PFAS Delay | Proprietary | 2.7 µm | N/A | + | 90m2/g | 2.0-9.0 | 60 °C (a) 40 °C (b) |
(a) Low pH; (b) High pH
The HALO® PFAS columns are available exclusively in Germany from MZ-Analysentechnik GmbH.
Representative MRM QA chromatogram showing the analysis of 17 PFAS compounds using HALO® PFAS.
PEAK IDENTITIES
1. PFBA 2. PFPeA 3. PFBS 4. PFHxA 5. PFHpA 6. PFHxS 7. PFOA 8. PFNA 9. PFOS 10. PFDA 11. PFDS 12. PFUnDA 13. PFDoDA 14. PFTrDA 15. PFTeDA 16. PFHxDA 17. PFOcDA
HALO PFAS Delay is an application assured column solution. The delay column is used to prevent background PFAS contamination from interfering with the PFAS of interest that are separated with the analytical column. The delay column bonded phase is a highly retentive endcapped silane chosen for its ability to demonstrate delay of background instrument PFAS contamination across multiple mobile phase conditions. For this reason, the HALO PFAS Delay column is placed upstream of the sample injector.
MZ-Analysentechnik GmbH is the official Advanced Materials Technology distributor in Germany. For further technical and price information please contact us!